Qualer Version 2.1 

Release Date: March 8th, 2017

Preview Environment Available: February 21st, 2017


A minor release will be made to the Qualer platform on March 8th, 2017.  This release will include various bug fixes, and added functionality to improve your overall experience with the Qualer system. A summary of the functionality included in this release, and an impact assessment for each feature is listed below. To learn more about each feature and see relevant screen captures, please click on the hyperlink in the table-- you will be redirected to the appropriate page. 


This release includes access to a client-specific preview environment to test functionality using your organization’s unique Qualer configuration.  This preview environment should be used by QA team and Qualer Customer Success Manager to update your Qualer Software Validation PQ as needed.  To access this preview environment simply use your browser to search for this domain format: [yourcompany].preview.qualer.com 


Note: Your Qualer preview site will require the same login credentials required for your production site, and any configuration updates or tests run in this preview environment will not impact or update your production site after the release. 


Feature ID
Risk Assessment
Application\ Module\ Function
Title
15589
Low
Qualer.Direct\Service\Work Orders\Direct Request
"Auto Tare" Command Enabled in Qualer.Direct Application
13957
Low
Qualer\Administration\Notification Setting
Notification Available After Client Requests "Local Pick-Up" in Service Request Form
16120
High
Qualer\Administration\Service Levels
Electronic Form Variations - Conversions (Service Measurement)
16053
Low
Qualer\Administration\Service Levels
Electronic Form Variations - Custom fields (Service Measurements)
15934
High
Qualer\Administration\Service Levels
Electronic Form Variations - Service Measurements
15732
Medium
Qualer\Administration\Service Workflows
Added Service Workflow Configuration
16095
Low
Qualer\Administration\Sites
Electronic Form Variations - TAR (Tool)
15782
Low
Qualer\Assets\Asset Manager\Edit
Additional Tool Management Functionality
15315
Low
Qualer\Assets\Maintenance Plans
Select Specific Weekdays Required for Maintenance Task Scheduling
15161
Low
Qualer\Assets\Service Requests
Search by Serial Number or Asset Tag in the "Pick From Assets" Pop-Up When Updating an Order
15808
High
Qualer\Bootstrap\Measurements
Electronic Form Variations - Custom fields (Measurements)
15810
High
Qualer\Bootstrap\Measurements
Electronic Form Variations - Reading entry logic
15895
High
Qualer\Bootstrap\Measurements
Electronic Form Variations - TAR (Measurements)
15871
Low
Qualer\Bootstrap\Measurements
Expanded PASS/FAIL Status and Criteria
15433
Medium
Qualer\Bootstrap\Measurements
Without a Specification Selected, Results Will Remain "N/A"
14976
Low
Qualer\Products\Manager
Enhanced Parts Module
15069
Low
Qualer\Service\Work items
Bulk Assign Technicians to Work Items
15702
Low
Qualer\Service\Work Orders
Automatic order creation changes
14714
Medium
Qualer\Service\Work Orders
Client Control of Sponsored Vendor Workflows
14823
Low
Qualer\Service\Work Orders
Work Order Navigation Improvements
15911
Low
Qualer\Service\Work Orders\Order
Add Ability to "Pass" or "Fail" an Order Through QC to Permissions List
14981
Low
Qualer\Service\Work Orders\Order
Display order item status in Service Scheduled grid
15833
Medium
Qualer\Service\Work Orders\Order\Price Tax
Hourly Charge Calculation in Fulfillment UI



Overview of Qualer's Risk Assessment Procedure


Risk is defined as Low, Moderate, or High based on the criteria identified in the following table: 

Consequence

Likelihood Based on the Frequency of Use


Low
(highly unlikely)

Medium

(credible/not likely)

High

(credible/likely)


Low
(Unregulated environment)

Low

Low or Medium

Medium

Medium
(Business operations)

Low or Medium

Medium

Medium or High

High
(IS0 9000/GxP environment)

Medium

Medium or High

High


Controls will be tested, as indicated:

Risk
Level of Testing
Low
Significant controls and functions will be tested for correct function. Invalid inputs and/or usage will not be tested, but will be noted if found.
Medium
All controls and functions will be tested for correct function. Invalid inputs and/or usage will not be tested, but will be noted if found.
High
All controls and functions will be tested for correct function. Where possible, invalid inputs and/or invalid usage will be tested to verify appropriate failure conditions.