Creating a Generic Specification
Step-by-Step Instructions:
- Hover over Products.
- Click on Product Specifications.
- Under the TOOLBOX section, a new action ‘Add Generic Specification’ is available. Once you click ‘Add Generic Specification’, a pop-up window will appear with a drop-down to select Product Type.
- Search through the category drop-downs to find the Product Type you would like to create a specification for. Once the Product Type has been selected, click ‘Continue’.
Enter the main title of the specification then a subtitle. Use the drop-downs on the right to specify if this is for a customer or a general specification that can be used for all customers (in this case, that means choosing Atommare Calibration). Then specify the group (ie Manufacturer, Customer, Handbook 44 spec). If the desired group is not there, click the pencil icon to create a new one. When finished, hit Save:
Now that the spec has been created, points can be added. First, choose if the spec should be force and/or fit:
FORCE: The number of points is driven by the specification. When a specification is selected, Qualer will match the Function with the Measurement name (in Service Levels). If it finds a match, Qualer will clear out the measurement lines and create a line for each nominal.
Description: If we have a single measurement parameter for temperature but we don’t know if the measurement is supposed to be in meters or inches, then when the specification is selected, the measurement will pull directly from the specification.
FIT: After a specification has been selected, Qualer will clear everything else from the form.
Description: If I am servicing a Centrifuge and the 3 primary tests are temperature, frequency, and time but when I show up at the job site the centrifuge doesn’t have a temperature control. Rather than having to adjust custom fields or SPEC itself. When the order is created it will have all measurements listed but if Qualer can’t locate all matching measurements then unmatched measurements will be removed.
Then click "Add new record":
Fill in the function name that matches the measurement parameter from the service level corresponding to this instrument type. Then name the specific test point being created by filling out "Specification." Next, choose the desired quantity/parameter and unit of measure:
You can choose from 6 different specification types: Percentage of Measured Value, Tolerance Range, Offset from Measured Value, Percentage + Count of Resolution, Parts Per Million of Measured Value, and Parts Per Million + Count of Resolution.
- Percentage of Measurement Value: See image above. Here, enter the value being measured (i.e. 100 grams) and then assign an accuracy the measurement will measure against (i.e. +/- 5%). This would mean any value between 95 grams to 105 grams is acceptable.
- Tolerance Range: To set up this type of specification, select a reference value (i.e. 100 grams) and now instead of choosing a percentage, tell the system a specific range that would be acceptable (i.e. 95-105).
- Percentage + Count of Resolution: To set up this type of specification, you would first choose a specification low limit (i.e. 0 grams) and a specification high limit (i.e. 100 grams). Next, enter the resolution of your equipment (i.e. 0.01 grams). Then, assign the acceptable tolerance percentage (i.e. 5%) and the count of resolution in this case, "2 counts" (this would mean your tolerance would be +/-5% +2 (measurement resolution).
- Parts Per Million of Measured Value: To set up this type of specification, select a reference value (i.e. 100 grams). Next, choose the accuracy in parts per million (i.e. 500) which would calculate to +/- 0.0005 grams.
- Parts Per Million + Count of Resolution: To set up this type of specification, choose a specification low limit (i.e. 0 grams) and a specification high limit (i.e. 100 grams). Next, enter the resolution of the equipment (i.e. 0.01 grams). Then, assign the acceptable tolerance in parts per million (i.e. 500) and the count of resolution in this case 2 (this would mean your tolerance would equal +/-0.0005 +2 (measurement resolution)).
Lastly, after configuring the specification point, select the green “OK”. Continue to add any additional test points. Finally, click Save.
To assign your newly created specification to an asset navigate to the desired asset within the asset manager/client asset manager / managed assets screens. Click on the asset serial number/asset tag/equipment ID to open up the 'Asset Info' screen. Click the blue 'Edit'